Nuclear Medicine Maintenance

Become a part of the BMET community & join our brand new forums to connect!

Maintenance, Daily Checks, and Troubleshooting of Nuclear Medicine Scanners

Proper maintenance and daily quality assurance (QA) checks are essential for the optimal performance, safety, and accuracy of nuclear medicine scanners, including SPECT (Single Photon Emission Computed Tomography) and PET (Positron Emission Tomography) systems. Regular maintenance prevents image artifacts, reduces downtime, and ensures compliance with regulatory standards (e.g., ACR, IAEA, FDA).


1. Daily Quality Control (QC) Checks

Daily QC tests ensure that the scanner is operating correctly and producing consistent, high-quality images. These checks are typically performed before patient scans and involve the following:

A. Gamma Camera (SPECT) Daily Checks

  1. Energy Window Calibration
    • Ensures the gamma camera is detecting photopeaks correctly (e.g., 140 keV for Tc-99m).
    • Uses a long-lived source (e.g., Co-57 flood source) for energy verification.
  2. Uniformity Check (Intrinsic & Extrinsic)
    • Uses a Co-57 flood phantom to verify detector uniformity.
    • Identifies non-uniformities caused by detector drift or crystal defects.
  3. Center of Rotation (COR) Verification
    • Ensures the camera’s rotational axis is properly aligned for accurate image reconstruction.

B. PET Scanner Daily Checks

  1. Blank Scan (Normalization Scan)
    • Uses a positron-emitting source (Ge-68 rod source) to check for detector uniformity.
    • Detects crystal malfunctions or PMT gain issues.
  2. Energy and Timing Calibration
    • Ensures accurate detection of 511 keV gamma photons from positron annihilation events.
    • Verifies coincidence timing window for PET’s Time-of-Flight (TOF) detection.
  3. Daily Sensitivity Check
    • Ensures the scanner detects the expected count rate per unit activity.
    • Identifies issues like block detector malfunctions or photomultiplier tube (PMT) failures.

2. Routine Preventive Maintenance

Preventive maintenance is performed weekly, monthly, or annually to prevent scanner failures and ensure regulatory compliance.

A. Weekly Checks

  • Collimator Inspection & Cleaning (SPECT)
    • Ensures lead collimators are free from dust or debris that could affect imaging.
  • Detector Linearity & Spatial Resolution Test
    • Uses a bar phantom to check spatial resolution & distortion.

B. Monthly & Quarterly Maintenance

  • High-Voltage Power Supply Check
    • Ensures power supply levels are stable, preventing PMT failures or gain drifts.
  • PMT Gain Adjustment (SPECT & PET)
    • Recalibrates PMTs or silicon photomultipliers (SiPMs) to maintain consistent light output.
  • Attenuation Correction Phantom Scan (PET/CT & SPECT/CT)
    • Uses a water-filled or uniform phantom to validate attenuation correction algorithms.

C. Annual Preventive Maintenance

Performed by biomedical engineers or service technicians, annual maintenance includes:

  1. Crystal Hydration Check (SPECT)
    • Checks for NaI(Tl) crystal moisture damage, which can degrade image quality.
  2. Detector Alignment & Gantry Calibration
    • Ensures gantry rotation & detector positioning remain precise.
  3. Cooling System & Fan Maintenance
    • PET and CT components require cooling system checks to prevent overheating.
  4. Software Updates & Workstation Optimization
    • Ensures the latest software patches and system optimizations are installed.

3. Common Issues and Troubleshooting

Despite regular maintenance, nuclear medicine scanners can experience mechanical, electronic, or software malfunctions. Below are common issues and their troubleshooting steps:

IssuePossible CauseTroubleshooting Steps
Non-Uniform Image (Hot/Cold Spots)Detector gain drift, faulty PMTs, crystal hydrationPerform flood field uniformity test and PMT recalibration.
Energy Window Shift (Incorrect Photopeak Detection)Voltage fluctuations, PMT agingRecalibrate energy window using Tc-99m or Co-57 source.
Ring or Arc Artifacts in PET ImagesDetector block failure, timing calibration errorPerform blank scan and replace defective detector block.
Low Sensitivity (Fewer Counts than Expected)Collimator blockage, contamination, PMT failureCheck collimator cleanliness, verify phantom test count rates.
Noisy Images or High Background ActivityContaminated dose, incorrect attenuation correctionCheck radiotracer purity, verify CT attenuation correction map.
SPECT COR MisalignmentGantry mechanical drift, detector imbalancePerform Center of Rotation (COR) correction.
PET Scanner Freezes or MalfunctionsSoftware bug, overheatingRestart system, check cooling fans & system logs.
CT Component of PET/CT or SPECT/CT Not FunctioningX-ray tube failure, power supply issueVerify tube lifespan, power connections, call vendor for service.

4. Radiation Safety & Quality Assurance

Since nuclear medicine scanners use ionizing radiation, strict radiation safety protocols must be followed:

A. Dose Calibration & Radiation Monitoring

  • Verify dose calibrator accuracy for correct radiopharmaceutical dosing.
  • Monitor room radiation levels using Geiger counters & dosimeters.

B. Radiation Shielding & Contamination Checks

  • Ensure lead shielding, collimators, and protective barriers are intact.
  • Regularly wipe-test surfaces to check for radioactive contamination.

C. Patient Safety Checks

  • Confirm correct radiotracer dose & patient identity before injection.
  • Educate patients on post-scan radiation precautions (e.g., avoiding close contact with infants).

5. Summary of Key Maintenance & Troubleshooting Practices

CategoryTasks PerformedFrequency
Daily QC ChecksEnergy window calibration, uniformity test, blank scan (PET)Before patient scanning
Weekly ChecksCollimator cleaning, linearity test, PMT gain checkWeekly
Monthly ChecksPower supply check, attenuation correction validationMonthly/Quarterly
Annual MaintenanceDetector realignment, CT calibration, cooling system serviceYearly
TroubleshootingImage artifacts, system errors, detector issuesAs needed
Radiation SafetyDose calibration, shielding integrity, contamination wipe testsRoutine

6. Conclusion

Regular maintenance, calibration, and troubleshooting of nuclear medicine scanners are crucial for ensuring accurate imaging, patient safety, and regulatory compliance. Daily checks focus on image quality assurance, while weekly, monthly, and annual maintenance prevent major system failures. Common troubleshooting techniques help address issues like image artifacts, detector failures, and radiation contamination. By implementing rigorous QA protocols, nuclear medicine departments can maintain optimal scanner performance, prolong equipment lifespan, and enhance diagnostic accuracy.